Vol. 42 (3): 546-549, May – June, 2016

doi: 10.1590/S1677-5538.IBJU.2015.0071


ORIGINAL ARTICLE

Dipstick Spot urine pH does not accurately represent 24 hour urine PH measured by an electrode

Mohamed Omar 1, Carl Sarkissian, Li Jianbo 1, Juan Calle 1, Manoj Monga 1
1 Glickman Urological & Kidney Institute – Cleveland Clinic Foundation, Cleveland, OH, USA

ABSTRACT      

Objectives: To determine whether spot urine pH measured by dipstick is an accurate representation of 24 hours urine pH measured by an electrode.
Materials and Methods: We retrospectively reviewed urine pH results of patients who presented to the urology stone clinic. For each patient we recorded the most recent pH result measured by dipstick from a spot urine sample that preceded the result of a 24-hour urine pH measured by the use of a pH electrode. Patients were excluded if there was a change in medications or dietary recommendations or if the two samples were more than 4 months apart. A difference of more than 0.5 pH was considered an inaccurate result.
Results: A total 600 patients were retrospectively reviewed for the pH results. The mean difference in pH between spot urine value and the 24 hours collection values was 0.52±0.45 pH. Higher pH was associated with lower accuracy (p<0.001). The accuracy of spot urine samples to predict 24-hour pH values of <5.5 was 68.9%, 68.2% for 5.5 to 6.5 and 35% for >6.5. Samples taken more than 75 days apart had only 49% the accuracy of more recent samples (p<0.002). The overall accuracy is lower than 80% (p<0.001). Influence of diurnal variation was not significant (p=0.588).
Conclusions: Spot urine pH by dipstick is not an accurate method for evaluation of the patients with urolithiasis. Patients with alkaline urine are more prone to error with reliance on spot urine pH.

Keywords: Urinary Tract; Electrodes; Urine